Revision
06/24/98
Press MODE to
Exit
<
to Resume
>
to Add New Line
+
to Insert
-
Delete
I
Figure
9
-
"Press
MODE to Exit"
Menu
The following functions are available on the "Press MODE to Exit" menu:
A.
Pressing the MODE key exits and returns to the "OTHER FUNCTIONS" menu.
B.
Pressing the left arrow key returns to the previous location on the AUTO SOLVENT
WASH menu.
C. Pressing the right arrow key adds a new line to the end of the program.
D. Pressing the PARAMETER SETTING plus
(+)
key inserts a new line before the current
line in the program.
E.
Pressing the PARAMETER SETTING minus
(-)
key deletes the current line in the
program.
9.
Press the right arrow to add a fourth line to the end of the Automated Solvent Wash program.
Enter the values listed in Table
4,
line
4
into the fourth line of the program. The parentheses
[
)]
next to line
4
indicates that this is the last line of the program.
10. Press the MODE key twice to reach the OTHER FUNCTIONS menu.
8.4
Wash Timing Params Menu
The Wash Time Parameter menu the provides control over the solvent delivery line purge time and
the pulse duration for pulsed draining.
8.4.1
Adjusting Wash Timing Params Procedure
1. Press the MODE key until the OTHER FUNCTIONS menu is displayed. Press the
PARAMETER SETTING minus key
(-)
to move the cursor to the Wash Time Params selection.
2.
Pressing the MODE key will display the Wash Timing Params menu as shown in Figure 10.
--WASH TIMING PARAMS-
Purge-----------
12.5
DrnPulse Closed-
2.0
Figure
10
-
WASH TIME
PARAMS
Menu
3.
To program the Purge time, use the left or right arrow key to move the cursor over the parameter.
Use the PARAMETER SETTING
(+
or
-)
to increment or decrement the purge time. The default
value for purge is 12.5 seconds.
4.
The Pulse duration for pulsed draining is defined by the amount of time the Lower Manifold
Valve is held open or closed. This can be varied by changing the values for the DrnPulse Open
and DrnPulse Closed parameters. To program either of the DrnPulse parameters use the left or
right arrow key to move the cursor over the desired parameter. Use the PARAMETER SETTING
(+
or
-)
to increment or decrement the selected parameters time. The default value for both the
DrnPulse Open and DrnPulse Closed parameters is 2 seconds.