Argonaut Technologies
Revision
06/24/98
--AUTO SOLVENT WASH-
Pgm:
1
5ml
Side:A&B
#
R
S
Add
Mix
Pul
1)
3xA
60
5:OO
25
Figure
8
-
AUTO SOLVENT WASH
Menu
-
first line displayed
A.
#:
Line number of the solvent wash. The default entry for this parameter is
1
representing
the first line of the program.
B.
R: Number of repetitive washes for the selected solvent. The default entry for this parameter
is
3
representing
3
repetitions of this line of the program.
C.
S: Solvent position selected. The default entry for this parameter is A representing the
Solvent in bottle A.
D.
Add: Solvent delivery time in seconds. The default entry for this parameter is 60
representing a delivery time of 60 seconds.
E. Mix: Agitation time in minutes and seconds. The default entry for this parameter is
5:00
representing an agitation time of
5
minutes and 0 seconds.
F. Pul: Pulsed draining time in seconds. The default entry for this parameter is 25 representing a
draining time of 25 seconds. If
Dm
is selected instead of Pul the default time would be 50.
Enter the values listed in Table
4,
line
1
into the first line of the Automated Solvent Wash
program. Use the right arrow soft key to align the cursor over the parameter to be changed and
press the PARAMETER SETTING (+or-) keys to incrementldecrement the value. *For entries
under the Add parameter use the measured delivery times for the respective solvents.
Table
4
-
Solvent Wash Program
-
Lines
1-4
After entering the Pul parameter press the right arrow key once to add a second line to the
Automated Solvent Wash program. The software creates a copy of the first line and enters it as
the next line of the program. Enter the values listed in Table
4,
line
2
into the second line of the
program.
After entering the Pul parameter press the right arrow key once to add a third line to the
Automated Solvent Wash program. The software creates
a
copy of the second line and enters it
as the next line of the program. Enter the values listed in Table
4,
line
3
into the third line of the
program. Another method to add or delete
a
line into the Automated Solvent Wash program is by
pressing the mode key. Press the mode key to display the menu shown in Figure