Argonaut Technologies Revision
06/24/98
8.5
Starting the Automated Solvent Wash Program
Before starting the program, Check that the Control Valves on side
B
of the Reactor Unit are turned
to CLOSED and CLOSED and that the MANIFOLD toggle valve is set to CLOSED. Rotate the
Reactor Unit to side A, set the Control Valves to SOLVENT and UTILITY
2
and the MANIFOLD
toggle valve to OPEN. Place the Glass Waste Tank below the Lower Manifold and open the Lower
Manifold Drain Valves. If there is liquid in the Reaction Vessels drain them before starting the
Automated Solvent Wash program.
Note: It is recommended to verify RV draining prior to running the automated solvent wash
program by manually filling the RVs with solvent followed by draining
Warning: Insure proper solvent deliver times before operating the Automated Solvent Wash
Option. Inappropriate solvent delivery times can cause over filling of the RVs creating cross
contamination between RVs, upper manifold clogs and solvent overflow into the ASWM.
Warning: Automated solvent washes generate a large amount of solvent waste. It is
recommended to connect a large secondary waste container to the Glass Waste Tank. Connect
a section of chemically resistant tubing to the valve on the Glass Waste Tank. The other end of
the tube should be connected to an appropriate secondary waste container.
8.5.1
Automated Solvent Wash Start Up Procedure
1.
Press the MODE key until the Agitation menu is displayed. Enter the Agitation Parameters to be
used during the Automated Solvent Wash program. For an explanation of agitation settings refer
to the Quest Users Manual. Once the program is started the only adjustment available for the
agitation parameters is the Agitation Needle Valve.
2.
Press the MODE key three times to display the AUTO SOLVENT WASH menu. A flashing
cursor is aligned to the right of Pgm: (program). Use the PARAMETER SETTING keys
(+
or
-)
to increment or decrement the program number until Pgm
1
is displayed.
3.
Press the AGITATION ON soft key to start the Automated Solvent Wash program. The program
will begin at line
1.
To start the Automated Solvent Wash program on a line other than line one, use the right or left arrow
keys to align the cursor over the desired line number. Pressing the AGITATION ON key displays the
menu shown in Figure
1
1.
Current Line
Top of
Pgm
'
<
'
Don't Start
Figure
11
-
"To Start At" Menu
A. To start at the Current Line press the MODE key.
B.
To start at the top of the program press the
(+)
PARAMETER key.
C. Pressing the left arrow key will not start the program, and will return to the previous
location on the AUTO SOLVENT WASH menu.